Advertisement

Experienced emergency medicine physicians cannot safely inflate or estimate endotracheal tube cuff pressure using standard techniques

      Abstract

      Objective

      Tracheal necrosis, stenosis, and rupture may result from overinflated endotracheal tube cuffs (ETTcs). We sought to determine the ability of faculty emergency medicine (EM) physicians to safely inflate ETTc as well as to estimate pressure of previously inflated ETTc.

      Methods

      Using a previously tested tracheal simulation model, we assessed EM physician inflation of ETTc pilot balloons. Participants also palpated the pilot balloon of 9 ETTc inflated to pressures ranging from extremely low to extremely high in a random order and reported their estimate of pressure.

      Results

      We sampled 41 faculty EM physicians from 5 EM residency programs. Using palpation, participants were only 22% sensitive detecting overinflated ETTc. The average ETTc pressure produced by inflation was more than 93 cm H2O (normal, 15-25 cm H2O).

      Conclusions

      Participants were unable to inflate ETTc to safe pressures or estimate pressure of ETTc by palpation. Clinicians should consider using devices to facilitate safe inflation and accurate measurement of ETTc pressure.
      To read this article in full you will need to make a payment
      One-time access price info
      • For academic or personal research use, select 'Academic and Personal'
      • For corporate R&D use, select 'Corporate R&D Professionals'

      Subscribe:

      Subscribe to The American Journal of Emergency Medicine
      Already a print subscriber? Claim online access
      Already an online subscriber? Sign in
      Institutional Access: Sign in to ScienceDirect

      References

        • Fan C.M.
        • Ko P.C.
        • Tsai K.
        • Chiang W.
        • Chang Y.
        • Chen W.
        • et al.
        Tracheal rupture complicating emergency endotracheal intubation.
        Am J Emerg Med. 2004; 22: 289-293
        • Harris R.
        • Joseph A.
        Acute tracheal rupture related to endotracheal intubation: case report.
        J Emerg Med. 2000; 18: 35-39
        • Hofmann H.S.
        • Rettig G.
        • Radke J.
        • Neef H.
        • Silber R.E.
        Iatrogenic rupture of the tracheobronchial tree.
        Eur J Cardiothorac Surg. 2002; 21: 649-652
        • Deslee G.
        • Brichet A.
        • Lebuffe G.
        • Copin M.C.
        • Ramon P.
        • Marquette C.H.
        Obstructive fibrinous tracheal pseudomembrane. A potentially fatal complication of tracheal intubation.
        Am J Respir Crit Care Med. 2000; 162: 1169-1171
        • Shelly W.M.
        • Dawson R.B.
        • May I.A.
        Cuffed tubes as a cause of tracheal stenosis.
        J Thorac Cardiovasc Surg. 1969; 57: 623-627
        • Weber A.L.
        • Grillo H.C.
        Tracheal stenosis, an analysis of 151 cases.
        Radiol Clin North Am. 1978; 16: 291-308
        • Otani S.
        • Fujii H.
        • Ishizu T.
        • et al.
        Recurrent nerve palsy after endotracheal intubation.
        Masui. 1998; 47: 350-355
        • Stewart S.
        • Secrest J.
        • Norwood B.
        • Zachary R.
        A comparison of endotracheal tube cuff pressures using estimation techniques and direct intracuff measurement.
        AANA J. 2003; 71: 443-447
        • Patel V.
        • Hahn I.H.
        • Preblick C.
        • Hoffman R.J.
        Graduating paramedic students are unable to estimate endotracheal tube cuff pressure by standard palpation technique [abstract].
        Prehosp Emerg Care. 2004; 8: 94
        • Guyton D.C.
        Endotracheal and tracheotomy tube cuff design: influence on tracheal damage.
        Crit Care Update. 1990; 1: 1-10
        • Somri M.
        • Fradis M.
        • Malatskey S.
        • Vaida S.
        • Gaitini L.
        Simple on-line endotracheal cuff pressure relief valve.
        Ann Otol Rhinol Laryngol. 2002; 111: 190-192
        • Guyton D.C.
        • Barlow M.R.
        • Besselievre T.R.
        Influence of airway pressure on minimum occlusive endotracheal tube cuff pressure.
        Crit Care Med. 1997; 25: 91-94
        • Curiel Garcia J.A.
        • Guerrero-Romero F.
        • Rodriguez-Moran M.
        Cuff pressure in endotracheal intubation: should it be routinely measured?.
        Gac Med Mex. 2001; 137: 179-182
        • Efferen L.S.
        • Elsakr A.
        Post-extubation stridor; risk factors and outcome.
        J Assoc Acad Minor Phys. 1998; 9: 65-68
        • Kastanos N.
        • Estopa M.R.
        • Marin Perez A.
        • Xaubet Mir A.
        • Agusti-Vidal A.
        Laryngotracheal injury due to endotracheal intubation: incidence, evolution, and predisposing factors. A prospective long-term study.
        Crit Care Med. 1983; 11: 362-367
        • Foroughi V.
        • Sripada R.
        Sensitivity of tactile examination of endotracheal tube intra-cuff pressure.
        Anesthesiology. 1997; 87 ([abstract]): 965
        • Ganner C.
        The accurate measurement of endotracheal tube cuff pressures.
        Br J Nurs. 2001; 10: 1127-1134
        • Fernandez R.
        • Blanch L.
        • Mancebo J.
        • Bonsoms N.
        • Artigas A.
        Endotracheal tube cuff pressure assessment: pitfalls of finger estimation and need for objective measurement.
        Crit Care Med. 1990; 18: 1423-1426
        • Braz J.R.
        • Navarro L.H.
        • Takata I.H.
        • Nascimento J.P.
        Endotracheal tube cuff pressure: need for precise measurement.
        Sao Paulo Med J. 1999; 117: 243-247
        • Farre R.
        • Rotger M.
        • Ferre M.
        • Torres A.
        • Navajas D.
        Automatic regulation of the cuff pressure in endotracheally-intubated patients.
        Eur Respir J. 2002; 20: 1010-1013
        • Komatsu H.
        • Mitsuhata H.
        • Hasegawa J.
        • Matsumoto S.
        Evaluation of a pressure and volume-relief instrument (modified Brandt's rediffusion system) to prevent increase in endotracheal tube cuff pressure.
        Masui. 1993; 42: 296-299
        • Vitkn S.A.
        • Lagasse R.S.
        • Poppers P.J.
        Application of the Grieshaber air system to maintain endotracheal tube cuff pressure.
        J Clin Anesth. 1990; 2: 45-47
        • Granja C.
        • Faraldo S.
        • Laguna P.
        • Gois L.
        Control of the endotracheal cuff balloon pressure as a method of preventing laryngotracheal lesions in critically ill intubated patients.