Eye injury from electrical weapon probes: Mechanisms and treatment

a b s t r a c t

Purpose: While generally reducing morbidity and mortality, TASER(R) electrical weapons have risks associated with their usage, including burn injuries and head and cervical trauma associated with uncontrolled falls. The pri- mary non-fatal complications appear to be significant eye injury but no analysis of the mechanisms or suggestED treatments has been published.

Methods: We used a biomechanical model to predict the risk of eye injury as a function of distance from the weapon muzzle to the eye. We compared our model results to recently published epidemiological findings. We also describe the typical presentation and suggest treatment options.

Results: The globe rupture model predicted that a globe rupture can be expected (50% risk) when the eye is within 6 m of the muzzle and decreases rapidly beyond that. This critical distance is 9 m for lens and retinal dam- age which is approximately the range of the most common probe cartridges. Beyond 9 m, hyphema is expected along with a perforation by the dart portion of the probe. Our prediction of globe rupture out to 6 m (out of a typ- ical range of 9 m) is consistent with the published risk of enucleation or unilateral blindness being 69 +- 18%, with an eye penetration.

Conclusions: Significant eye injury is expected from a penetration by an electrical weapon probe at close range. The risk decreases rapidly at extended distances from the muzzle. Not all penetrating globe injuries from electri- cal weapon probes will result in blindness.

(C) 2018

Introduction

Electronic control with the CEW (conducted electrical weapon) has gained widespread acceptance with law-enforcement, as the preferred less-Lethal force option due to its dramatic injury reduction compared to other control tools. Large prospective studies have found reductions in the subject injury rate of about 65% [1]. Of the 310,000 annual CEW field uses, 1 in 3500 is involved in a non-firearm ARD (arrest-related death) vs. the baseline ARD rate of 1:1000 [2]. This is consistent with a 2/3 reduction in fatal police shootings where CEW usage is not overly restricted [3].

Electrical weapons are, after all, weapons, and there are indeed risks associated with their usage, including fatalities from falls and burns [4-

* Corresponding author at: Box 23, Crystal Bay, MN 55323, United States.

E-mail address: [email protected] (M.W. Kroll).

6]. They launch probes with darts and hence there is a risk of significant eye injury. See Fig. 1. The risk of penetrating eye injury has been recently reported as 1:123,000 confidence limits [85,000, 178,000] by Wilson score interval [7]. The goal of this paper is to analyze the risks of such in- jury from an analytical framework and to provide guidelines for emer- gent care.

A CEW has both laser and fixed sights. The X26E CEW series has a sin- gle laser that approximately aligns with the top dart. The lower probe is launched at a separation angle of 8? below the laser line. Newer models such as the X2 have dual laser sights and a separation angle of 7?. To ob- tain a high level of motor-nerve mediated neuromuscular incapacita- tion — on the front of the body — there must be a probe separation of at least 30 cm (12 in) [8]. This requirement, of a large spread, adds to a risk of a facial impact — especially if the subject ducks to avoid the probes. The critical range for a devastating eye injury such as a globe rupture has not been established. The optimal therapy for a CEW probe injury

has also not been published.

https://doi.org/10.1016/j.ajem.2018.06.004

0735-6757/(C) 2018

Fig. 1. X26(E) CEW during Probe deployment.

Methods

The prediction of eye injuries from ballistic impact has been studied by: (1) finite-element simulations that investigate tissue-level mechan- ical response from an impact, and (2) empirical models that correlate injury outcome to projectile ballistics. The simulations have proven use- ful in the investigation of potential mechanisms for different injury out- comes [9-11]. The empirical models have highlighted the significance of 3 parameters for eye injury risk: (1) projectile mass, (2) projectile veloc- ity, and (3) projectile size [12-15]. These 3 parameters can be combined to simply calculate the energy density or ANE (area-normalized energy) in kJ/m2, which is the kinetic energy of the projectile divided by the projectile’s cross-sectional area [15]. The ANE was shown to predict eye injury with logistic regression analysis [12,14].

Comprehensive risk functions have been derived from experimental eye impact tests using post-mortem human eyes, live animal surro- gates, and post-mortem animal eyes [9,11-17]. The predictions, using a logistic transformation of the ANE, have been validated for recrea- tional weapons such as airsoft pellet and paintball guns [9,13]. These risk functions have also been validated over a broad range of small pro- jectiles of various velocities, and have been compared against predicted corneo-scleral stresses and internal eye pressure from matched compu- tational simulations [11,16].

These risk functions have been based on spherical and blunt cylin- drical objects ranging in diameter from 4.4 mm to 19.9 mm. The CEW probe diameter (5.5 mm) falls in between BBs (4.4 mm) and airsoft pel- lets (6.0 mm). Overall, the energy profiles of the objects were very sim- ilar. The BB velocities range from 115 to 145 m/s and the ballistic energy ranges from 2.2-3.6 J, compared to the 2.6 J of the CEW probe at the muzzle. The ANE of BBs ranges from 145 to 235 kJ/m2, compared to the 112 kJ/m2 initial value for the CEW probe.

The dart (tip) of the probe will markedly increase the chance of a penetrating eye injury, therefore, the risk functions are not to be interpreted as an indicator of dart induced injuries (e.g. they cannot predict the risk of Penetrating injuries as a result of a direct hit by the probe-tip per se). For the CEW probe, we applied these injury func- tions by assuming that the sharp “dart” portion (of the probe) will al- ways penetrate the globe and hence that is not part of the modeling (see Fig. 2). Depending on where the dart penetrates, the injuries (from the dart portion, per se) are typically pin-hole defects or blur- ring. There have been cases of globe rupture where the dart-tip

lodged in posterior structures and thus presumably effected other damage.

The models predict 4 different types of eye injury: (1) hyphema,

(2) retinal damage, (3) lens damage, and (4) globe rupture. This model- ing is based only on the impact of the shoulders of the main part of the probe. Hyphema will occur first with a centered impact. With greater impact energy, this is followed by both lens and retinal damage which have similar thresholds. The lens damage is due to the local mechanical stresses. The retinal damage is hypothesized to be due to: (1) high globe deformation and localized stresses causing tissue failure, or (2) contre- coup injury due to pressure shock transmission. Finally, the globe rup- ture occurs when tensile stress thresholds (particularly in the corneo- scleral shell) are exceeded.

The X26E CEW probe is shown in Fig. 2. The probe ballistics (mass 2.85 g, diameter 5.4 mm), were utilized to calculate the kinetic energy and density (ANE) from the muzzle (0 m distance) to 10.67 m. The ini- tial probe velocity and energy of 42.7 m/s (2.6 J) decreases to 10.7 m/s (0.2 J) at ~11 m. The ANE (kJ/m2) was utilized to calculate the risk of eye injury (hyphema, retinal damage, lens damage, and globe rupture) based on the logistic injury risk functions of Duma and Kennedy [12,14]. The ANE, relative to flight distance, is plotted in Fig. 3.

Fig. 2. Single-shot probe as used in X26 models. Dart portion is 13 mm long; other version has 9 mm dart.

Fig. 3. Area-normalized energy (energy density) of an X26E CEW standard probe.

Results

This area-normalized energy was used to calculate the risk of the 4 eye injuries as a function of distance as shown in Fig. 4. These risk func- tions predict a 50% risk of a globe rupture at distances b6.2 m. The threshold distance for retinal and lens damage is 8.6 m; for hyphema the threshold is 9.6 m. With a preferred muzzle-to-subject targeting range of 2.1-4.6 m (7-15 ft.) each of these thresholds is within the pre- ferred range and hence a subset of these injuries are expected in a typ- ical probe deployment with eye contact.

It should be noted that these injury-risk curves represent the risk of a “perfectly” errant probe deployment, where the eye is struck directly at a perpendicular angle in the center and there is no eyelid absorbing probe energy. The modeling — and all of the cases listed below — is for the M26 and X26E models. Newer models (e.g. X2 and T7) have dif- ferent probes.

The effects of longer-distance (lower impact energy) probe landings are significantly different. In that case, the “specific energy” is lower so that the impact of the probe “shoulder” is not the primary factor in the eye injury (along with hyphema from the impact of the probe shoul- der). Depending on the landing location, the sharp “dart” portion will penetrate and may cause only acute damage, blurring, or a pinhole de- fect. However, if the dart penetrates the lens regardless of shoulder im- pact then we would expect more clinically significant injury than, say, mere hyphema.

Fig. 4. Risk of various eye injuries by Dumas-Kennedy model as function of distance.

This is well illustrated by a published case report of a penetrating dart injury apparently without significant shoulder impact [18]. The dart entered the temporal limbus, traversed the inferotemporal iris and lens, and exited the inferior pars plana. A vitreous hemorrhage was present. Nylon sutures were used to close the corneal and scleral defects. The patient regained best-corrected visual acuity of 6/18.

It must also be stressed an eyelid appears to offer significant protec- tion to the damage from the collar impact. A published case report de- scribes a patient in which the probe landed with the eyelids closed [19]. The dart penetrated just above the infraorbital bony margin. There was a small subconjunctival inferior hemorrhage, but the cornea was intact. After dart removal, the wound was closed with vicryl su- tures. The patient ended up with a pinhole visual acuity of 6/9.

Discussion

We believe that this paper represents the first biomechanical analy- sis of the risk of penetrating eye injury from electrical weapon probes. We applied a validated biomechanical eye injury model which shows that we can expect (1) hyphema, (2) lens damage, (3) retinal damage, and (4) globe rupture with existing probes at decreasing deployment distances for a centered impact.

These findings appear to comport with published data finding 12 enucleations and 6 cases of complete blindness and thus the majority (18/28) of probe impacts resulted in a loss of vision [7]. In total there have been 7 cases of partial blindness, and 2 cases of normal vision after successful Surgical repair. (There was a case lost to follow-up after a surgical repair attempt.) See Table 1 for details. With 18 identi- fied cases, there is a risk of 0.64, CI [0.47-0.82] for unilateral blindness or enucleation from a penetrating eye injury, primarily from globe rup- ture. See Fig. 5.

Globe rupture was usually associated with blindness or enucleation. However, there were 4 cases of unilateral blindness where the records were insufficient for us to determine if there was a globe rupture. Dart penetration (without significant impact energy from the probe

Table 1

Penetrating globe injury field use cases.

Year Age Sex State Code Source

2004

50

M

AU

N

Ng & Chehade [19]

2005

21

M

CA

P

Kroll et al. [7]

2005

21

M

OR

P

Chen et al. [18]

2007

27

M

TX

B

Kroll et al. [7]

2007

18

M

CA

P

Kroll et al. [7]

2008

24

M

CA

B

Kroll et al. [7]

2008

25

M

CN

N

Han et al. [35]

2008

46

M

KR

E

Park et al. [36]

2009

43

M

TN

E

Kroll et al. [7]

2009

26

M

MO

E

Teymoorian [37]

2011

52

M

PA

E

Kroll et al. [7]

2011

19

M

TX

B

Kroll et al. [7]

2011

34

M

NM

P

Kroll et al. [7]

2011

27

M

FL

P

Kroll et al. [7]

2012

47

F

TX

UNK

Li and Hamill [38]

2013

24

M

OH

P

Kroll et al. [7]

2013

22

M

FL

B

Kroll et al. [7]

2014

36

F

QU

E

Kroll et al. [7]

2014

55

M

Sctl

E

Kroll et al. [7]

2014

31

M

CA

E

Kroll et al. [7]

2014

21

M

WA

B

Kroll et al. [7]

2014

24

M

NY

E

Rafailov et al. [39]

2014

36

M

TX

E

Cahill and Jardeleza [40]

2015

22

F

MO

P

Kroll et al. [7]

2015

26

M

AL

E

Kroll et al. [7]

2015

20

M

QU

B

Kroll et al. [7]

2015

53

M

FL

E

Kroll et al. [7]

2017

31

M

NY

E

Kroll et al. [7]

Sctl = Scotland. QU = Queensland, Australia.

P = partial blindness (pinhole or Blurred vision), B = blindness, E = enucleation.

Fig. 5. Outcomes, including those unknown (UNK), after penetrating globe injury.

shoulder) was typically only associated with pinhole defects or minor blurring.

Electroporation is another possible means of tissue injury from the voltage spikes of CEWs. This has been well studied and the effects are limited to b1 mm from the probes where it helps sterilize the probes [20,21]. We suspect that there may have been some confusion caused by the media myth that CEWs deliver 50,000 V to the body vs. the actual 600 V [22].

We remain open-minded but somewhat skeptical of the occasional hypothesis of non-penetrating electrical eye injuries from electrical weapons for 2 reasons. (1) The electrical output (b2 W) appears too low to cause such damage by thermal effects, and (2) cases such as that of Jey, where there was a peri-ocular probe yet no reduction in vi- sion [23].

The Axon-competitor models of probe-capable CEWs, PhaZZer(TM) En- forcer and the Condor(TM) Spark use essentially identical probes, and thus a similar eye injury profile is expected.

Within the limits of the published experience, the biomechanical law enforcement and pa”>model predictions appear to be consistent with field results.

There is a case of possible mechanical injury from the probe impacting near the eye. An Ontario, Canada provincial police officer was responding to a call about an out-of-control man at a home where he deployed his CEW and a probe lodged near the man’s left eye- lid. The man was taken to a local nursing station, treated, and released, but returned the next day with pain and swelling to his eye, and later underwent surgery for a detached retina.

We found another eyelid landing case that did not, however, pro- duce ocular injury. A 36-year-old man was wanted for burglary, auto theft, and parole violations in New York. Deputies found him hiding in a field. When they approached him, the man pulled a knife. A deputy de- ployed his CEW and a probe struck the man in the upper left eyelid. The man apparently suffered no long-term injury or loss of vision.

Jey reported on a California subject struck in the left orbit [23]. CT findings revealed left preseptal hematoma, a small amount of hema- toma under the left eyelid, mild proptosis, and no discernable evidence of orbital fracture, optic nerve, or ocular findings. Emergent evaluation under anesthesia noted an injury of the eyelid, with disruption of the globe in the posterior location with partial extrusion of blood and vitre- ous. He was treated for vitreous hemorrhage and a retinal tear, which

necessitated retinal photocoagulation. On the patient’s last visit, he had 20/20 vision in his eye.

Sayegh reported a Maryland subject with a probe embedded in his right lower eyelid [24]. The probe created a subRetinal hemorrhage that progressed into an exudative retinal detachment over 3 days. The retinal detachment gradually resolved, and visual acuity improved over the next 2 months. The report suggested that the injuries were electrical in nature. However, the authors supported that hypothesis with citations to lightning injuries and appeared to be unaware of the low 2-watt output of CEWs. It is possible that there was some electropo- ration damage [25,26].

In our search, we found another example of alleged electrical (non- ballistic) injury. Officers, in Florida, were attempting to subdue a bur- glary suspect. The suspect managed to turn the CEW toward a female officer’s face just before it deployed. A probe hit her in the right temple; the other hit her near the bridge of her nose, just below her right eye- brow. According to media reports, it formed a direct current through her right eye, which allegedly left her blinded. We consider this possibil- ity unlikely, given the significant difference in power between the CEW and RF cauterization devices. The low (b2 W) output of TASER CEWs can be compared to the typical 50-100 W required for electrocautery.

A final example of alleged electrical-ocular injury was the case re- ported by Seth [27]. A 35-year-old male presented, in Connecticut, with traumatic iritis, angle-recession glaucoma, and retinal dialysis sec- ondary to “blunt” trauma from a CEW probe in his right eye 6 days pre- vious. He was said to have an “electrical” cataract in the left eye, and minor skin burns on the left eyelid. The ophthalmologist successfully repaired the right eye, but the patient was lost to follow-up before he could attempt to repair the left eye. Cataract formation may occur in up to 5% of electrical injuries [28]. However, this usually involves mas- sive energies, significantly greater than CEWs deliver, high enough to cause loss of consciousness and massive tissue necrosis [29,30]. For ex- ample, electrically-induced cataracts have been associated with extrem- ity amputations [31]. Electrically-induced cataracts typically take months to develop but may occur within hours in cases of severe burns close to the eye [29,30].

There is no obvious lesson for the user of an electrical weapon. Often, a law-enforcement officer does not have control over the range for a frontal confrontation. Even if the range was the officer’s choice there is no clear optimal range. A range of 6-9 m significantly decreases the risk of globe rupture in the event of a globe penetration. However, at, say, 7 m the aiming accuracy is proportionately reduced. With the probe dispersion angle of 8? the probe spread is 1 m so it is difficult to place the bottom probe in a leg while not having the top probe contact the face.

Guidelines for law enforcement and paramedics

When First responders confirm an embedded CEW probe in the periocular or ocular tissues they should follow general principles for emergency management of embedded foreign bodies. Namely, the probe should be left in place — as is — without attempting to remove or alter its position in any way as this could produce more harm to the visual structures.

Snip away any wire still attached to the probe. Any safe means of protecting the probe from changes in position may be attempted such as affixing a truncated paper cup over the area to act as a shield. Do not attempt to secure the probe to the patient’s face or eyelid as eye movements may then cause a probe position change and further injury. The patient should be transported to an emergency department imme- diately where ophthalmologists should be consulted.

Treatment of probe-induced eye injury

We offer the guidance below for treating a CEW probe penetrating injury. This is based on our (RS & NKS) specific experience [32]. It is

also based on the present gold standard management of ruptured globes after injury [33,34].

In the event of a probe injury to the eye, it is imperative (1) to have only ophthalmological surgeons remove the probe, and (2) that care be taken to rule out a ruptured globe. Patients with anterior globe rupture often present with obvious distortion of normal ophthalmic anatomy; typically, these injuries are corneal or scleral lacerations which are asso- ciated with extrusion of intraocular contents. It is common to see iris prolapse from corneal wounds or uveal prolapse from scleral wounds. In the latter scenario, uvea, a dark material, which is often attached to vitreous (a clear gel) can be seen emanating from a scleral laceration. Anterior globe rupture can often be identified using a penlight.

In some cases, anterior globe rupture is less obvious, and examina-

tion with a slit lamp is necessary. In order to evaluate a conspicuous an- terior globe rupture, an ophthalmologist should be expeditiously consulted to perform a Seidel test. In this examination, the surface of the eye is anesthetized with topical tetracaine or proparacaine drops, and a fluorescein strip is gently brushed across the site of suspected in- jury. In the case of a globe rupture, fluorescein-stained material will em- anate from the injury in the form of a leak. Posterior globe rupture (much less likely) is more difficult to evaluate as it may not be obvious on initial exam. In some cases, the eye may appear shrunken due to loss of vitreous from an unidentified site. If the eye appears normal on ante- rior exam but posterior rupture is suspected, imaging should be per- formed. CT scan of the brain and orbits without contrast (preferably 1 mm slices) is the most efficient and safe Imaging study. MRI is contra- indicated in the presence of metallic foreign bodies. Alternatively, gentle ultrasound can be used to identify intraocular foreign bodies or poste- rior globe rupture. Ultrasonography should not be performed in the set- ting of concurrent anterior globe rupture.

Once globe rupture is determined, further examination of the globe should be deferred until surgical repair. No pressure should be placed on the globe and the eye should be protected with a shield. A patch should not be placed over the eye.

CEW probes have a 5% rate of Staphylococcus aureus contamination [21]. Systemic antibiotics with broad coverage such as Vancomycin and Cefazolin or Ciprofloxacin or 4th generation quinolones (moxifloxacin or gatifloxacin for better intraocular penetration) should be administered. Children with ruptured globes can be given Cefazolin and Gentamycin.

Endophthalmitis, ocular inflammation often in the setting of infec- tion, is more likely in cases with dirty injuries, retained intraocular for- eign bodies, rupture of the lens capsule, or long delays until surgical repair. Pain management and anti-emetic medications should also be administered as needed.

Surgical repair by a trained ophthalmologist should be arranged as soon as possible. Some practitioners recommend the use of intravitreal antibiotics during globe repair but this is not currently the standard of care. If the globe rupture is associated with skin lacerations near the eye, the laceration repair should be deferred in order to avoid accidental manual pressure on the globe during suturing. In cases where the globe does not appear to be salvageable by surgical intervention, enucleation of the globe should be considered. These surgeries should be performed within 7-14 days of the initial trauma to prevent the Rare occurrence of sympathetic ophthalmia – an autoimmune inflammatory attack on the contralateral eye that can compromise vision of that globe.

Limitations

This work was based on a biomechanical model which was primarily validated with spherical and cylindrical projectiles. It is possible that the cylindrical CEW probe collar has a different eye injury profile when pre- ceded by the sharp dart portion. A prospective experimental study could generate superior data compared to our modeling results. How- ever, a relevant experiment would have difficulty obtaining ethical approvals.

Our modeling results appear to be consistent with published epide- miological data. However, those data are based largely on secondary sources.

Conclusions

A biomechanical model predicts that a globe rupture can be ex- pected (p = .5) when the eye is within 6 m of the CEW cartridge muzzle and decreases with an increasing distance. Significant eye injury is ex- pected from a penetration by an electrical weapon probe at close range. Not all penetrating globe injuries from electrical weapon probes will result in blindness.

Disclosures

MWK is a member of the corporate and scientific advisory boards of AXON (fka TASER) who partially funded this work. MAB is the principal of LAAW, LLC and a former employee of Axon. HEW is a retired police chief and received a speaking honorarium from Axon. MWK, MAB, and HEW have done expert witnessing in use-of-force cases. MBR, EAK, NKS, and RS report no conflicts.

References

  1. MacDonald JM, Kaminski RJ, Smith MR. The effect of less-lethal weapons on injuries in police use-of-force events. Am J Public Health 2009;99(12):2268-74.
  2. Kroll MW, Lakkireddy DR, Stone JR, Luceri RM. TASER electronic control devices and cardiac arrests: coincidental or causal? Circulation 2014;129(1):93-100.
  3. Ferdik FV, Kaminski RJ, Cooney MD, Sevigny EL. The influence of agency policies on conducted energy device use and police use of lethal force. Police Q 2014;17.4: 328-58.
  4. Kroll MW, Adamec J, Wetli CV, Williams HE. Fatal traumatic brain injury with elec- trical weapon falls. J Forensic Leg Med 2016;43:12-9.
  5. Clarke C, Andrews SP. The ignitability of petrol vapours and potential for vapour phase explosion by use of TASER(R) law enforcement electronic control device. Sci Justice 2014;54(6):412-20.
  6. Kroll M, Ritter M, Williams H. Fatal and non-fatal burn injuries with electrical weapons and explosive fumes. J Forensic Leg Med 2017;50:6-11.
  7. Kroll MW, Ritter MB, Kennedy EA, et al. Eye injuries from electrical weapon probes: incidents, prevalence, and legal implications. J Forensic Leg Med 2018;55:52-7.
  8. Ho J, Dawes D, Miner J, Kunz S, Nelson R, Sweeney J. Conducted electrical weapon incapacitation during a goal-directed task as a function of probe spread. Forensic Sci Med Pathol 2012;8(4):358-66.
  9. Kennedy EA, Stitzel JD, Duma SM. Matched experimental and computational simu- lations of paintball eye impacts. Biomed Sci Instrum 2007;44:243-8.
  10. Stitzel JD, Duma SM, Cormier JM, Herring IP. A nonlinear finite element model of the eye with experimental validation for the prediction of globe rupture. Paper pre- sented at: SAE conference proceedings; 2002.
  11. Weaver AA, Kennedy EA, Duma SM, Stitzel JD. Evaluation of different projectiles in matched experimental eye impact simulations. J Biomech Eng 2011;133(3):031002.
  12. Duma SM, Ng TP, Kennedy EA, Stitzel JD, Herring IP, Kuhn F. Determination of signif- icant parameters for eye injury risk from projectiles. J Trauma Acute Care Surg 2005; 59(4):960-4.
  13. Kennedy EA, Ng TP, Duma SM. Evaluating eye injury risk of airsoft pellet guns by parametric risk functions. Biomed Sci Instrum 2005;42:7-12.
  14. Kennedy EA, Ng TP, McNally C, Stitzel JD, Duma SM. Risk functions for human and porcine eye rupture based on projectile characteristics of blunt objects. Stapp Car Crash J 2006;50:651.
  15. Kennedy E, Duma S. FINAL REPORT eye injury risk functions for human and FOCUS eyes: hyphema, lens dislocation, and retinal damage. US Army Medical Research and Materiel Command/US Army Aeromedical Research Laboratory; 2010.
  16. Stitzel Jr JD, Weaver AA, Kennedy EA, Duma SM. Evaluation of different projectiles in matched experimental eye impact simulations. Invest Ophthalmol Vis Sci 2011;52 (14):5583.
  17. Brozoski F, Kennedy E, Duma S, Street S. FINAL REPORT eye injury risk functions for human and FOCUS eyes: hyphema, lens dislocation, and retinal damage. Update; 2011.
  18. Chen SL, Richard CK, Murthy RC, Lauer AK. Perforating ocular injury by TASER(R). Clin

    Experiment Ophthalmol 2006;34(4):378-80.

    Ng W, Chehade M. Taser penetrating ocular injury. Am J Ophthalmol 2005;139(4): 713-5.

  19. Gowrishankar TR, Esser AT, Smith KC, Burns SK, Weaver JC. In silico estimates of cell electroporation by electrical incapacitation waveforms. Conf Proc IEEE EMBC 2009; 2009:6505-8.
  20. Kroll MW, Ritter MB, Guilbault RA, Panescu D. Infection risk from conducted electri- cal weapon probes: what do we know? J Forensic Sci 2016;61(6):1556-62.
  21. Dawes DM, Ho JD, Kroll MW, Miner JR. Electrical characteristics of an electronic con- trol device under a physiologic load: a brief report. Pacing Clin Electrophysiol 2010; 33(3):330-6.
  22. Jey A, Hull P, Kravchuk V, Carillo B, Martel JB. Emergent diagnosis and management of TASER penetrating ocular injury. Am J Emerg Med 2016;34(8):1740e1743-1745.
  23. Sayegh RR, Madsen KA, Adler JD, Johnson MA, Mathews MK. Diffuse retinal injury from a non-penetrating TASER dart. Doc Ophthalmol 2011;123.2:135.
  24. Kroll MW, Dawes DM, Heegaard WG. TASER electronic control devices and eye inju- ries. Doc Ophthalmol 2012;124(2):157-9.
  25. Sayegh RR, Madsen KA, Adler JD, Johnson MA, Mathews MK. Response to TASER electronic control devices and eye injuries. Doc Ophthalmol 2012;124(2):161-2.
  26. Seth RK, Abedi G, Daccache AJ, Tsai JC. Cataract secondary to Electrical shock from a Taser gun. J Cataract Refract Surg 2007;33(9):1664-5.
  27. Martinez JA, Nguyen T. Electrical injuries. South Med J 2000;93(12):1165-8.
  28. Al Rabiah SM, Archer DB, Millar R, Collins AD, Shepherd WF. Electrical injury of the eye. Int Ophthalmol 1987;11(1):31-40.
  29. Boozalis GT, Purdue GF, Hunt JL, McCulley JP. Ocular changes from electrical burn in- juries. A literature review and report of cases. J Burn Care Rehabil 1991;12(5): 458-62.
  30. Long JC. Electric cataract: report of three cases. Am J Ophthalmol 1966;61(5 Pt 2): 1235-9.
  31. Rafailov L, Temnogorod J, Tsai FF, Shinder R. Impaled orbital TASER probe injury re- quiring primary enucleation. Ophthal Plast Reconstr Surg 2017;33.3S:S176-7.
  32. Scott R. The injured eye. Philos Trans R Soc Lond B Biol Sci 2011;366(1562):251-60.
  33. Li X, Zarbin MA, Bhagat N. Pediatric Open globe injury: a review of the literature. J Emerg Trauma Shock 2015;8(4):216-23.
  34. Han JS, Chopra A, Carr D. Ophthalmic injuries from a TASER. CJEM 2009;11(1):90-3.
  35. Park KH, Kim WJ, Kang YJ, Park JO, Kang W, Lee SG. A case of penetrating eye injury induced by Taser gun. J Korean Soc Emerg Med 2009;20(5):590-2.
  36. Teymoorian S, San Filippo AN, Poulose AK, Lyon DB. Perforating globe injury from Taser trauma. Ophthal Plast Reconstr Surg 2010;26(4):306-8.
  37. Li JY, Hamill MB. Catastrophic globe disruption as a result of a TASER injury. J Emerg Med 2013;44(1):65-7.
  38. Rafailov L, Temnogorod J, Tsai FF, Shinder R. Impaled orbital TASER probe injury re- quiring primary enucleation. Ophthal Plast Reconstr Surg 2017;33(3S Suppl 1): S176-7.
  39. Cahill C, Jardeleza M. Penetrating ocular injury from Taser. Austin J Clin Ophthalmol 2015;2(2):1-2.